ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.
FDA Device Recall #Z-0723-2017 — Class II — October 21, 2016
Recall Summary
| Recall Number | Z-0723-2017 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 21, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Healthcare Diagnostics, Inc. |
| Location | Tarrytown, NY |
| Product Type | Devices |
| Quantity | US: 23 systems; Foreign: 141 systems |
Product Description
ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.
Reason for Recall
Siemens is investigating an issue which may cause an incorrect result to be reported. Siemens received two reports of multiple discordant records for the same Sample ID in the ADVIA 560 Hematology System database that occurred during the installation of the systems. The database should only contain one record of a Sample ID number for any given time and date. If there are multiple records for the same Sample ID, it is possible that multiple results may be manually or automatically sent to the Laboratory Information System (LIS), printed or displayed on the results report screen.
Distribution Pattern
Worldwide Distribution - US Distribution to the states of including Puerto Rico and : CA, CO, FL, IL, MI, NJ, NY, OH, TX, WA, WI, WI and WY., and to the countries of : Angola, Australia, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, France, Germany, Greece, India, Italy, Japan, Kenya, Malaysia, Mexico, Myanmar, Netherlands, Norway, Pakistan, Portugal, Saudi Arabia, Slovenia, Spain, Switzerland, Turkey, Uganda, United Kingdom, Vietnam.
Lot / Code Information
All serial numbers of ADVIA 560 Hematology Systems
Other Recalls from Siemens Healthcare Diagnostics, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1484-2026 | Class II | Atellica CH Urine Albumin (UAlb). Material Numb... | Jan 20, 2026 |
| Z-0491-2026 | Class II | IMMULITE 2000 Oak Mix Specific Allergen; Sieme... | Oct 6, 2025 |
| Z-0492-2026 | Class II | 3gAllergy Specific IgE Universal Kit; Kit Comp... | Oct 6, 2025 |
| Z-0493-2026 | Class II | Food Panel 3 Specific Allergen; Siemens Materi... | Oct 6, 2025 |
| Z-0366-2026 | Class II | Assay: IMMULITE 2000 Intact PTH; Test Code: iP... | Sep 22, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.