EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE¿ device is a user controlled, ste...

FDA Device Recall #Z-1927-2025 — Class II — May 23, 2025

Recall Summary

Recall Number Z-1927-2025
Classification Class II — Moderate risk
Date Initiated May 23, 2025
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm QUANTUM SURGICAL SAS
Location montpellier
Product Type Devices
Quantity 4 units in U.S.

Product Description

EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE¿ device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G.

Reason for Recall

Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user.

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution in the states of FL,IL, IN, and TX. The countries of France, Germany and United Kingdom.

Lot / Code Information

Model Number: 30-0001 UDI-DI code: 03760305400031 Software Version: 1.0.6.1 Serial Numbers (U.S.): 0322002, 0323006, 0323003 and 0324001

Other Recalls from QUANTUM SURGICAL SAS

Recall # Classification Product Date
Z-0037-2025 Class II The EPIONE device v1.0.2 is a user controlled, ... Aug 5, 2024

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.