Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF ...
FDA Device Recall #Z-3240-2018 — Class II — July 26, 2018
Recall Summary
| Recall Number | Z-3240-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 26, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Stryker GmbH |
| Location | Selzach |
| Product Type | Devices |
| Quantity | 233 |
Product Description
Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
Reason for Recall
When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.
Distribution Pattern
Nationwide
Lot / Code Information
Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
Other Recalls from Stryker GmbH
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1914-2025 | Class II | Long Nail, RIGHT Gamma4 10x320mm x 125, compone... | May 28, 2025 |
| Z-1940-2025 | Class II | Femoral Nail, RIGHT T2 GTN 8x300 mm. Catalog Nu... | May 22, 2025 |
| Z-1939-2025 | Class II | Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Num... | May 22, 2025 |
| Z-1620-2025 | Class II | Guide Wire with Ruler Tube 3x800 mm DIA; Catalo... | Mar 28, 2025 |
| Z-1621-2025 | Class II | Guide Wire with Ruler Tube 3x1000 mm DIA; Catal... | Mar 28, 2025 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.