Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets; Paradigm Silhouette infusion se...

FDA Device Recall #Z-0524-2015 — Class II — November 4, 2014

Recall Summary

Recall Number Z-0524-2015
Classification Class II — Moderate risk
Date Initiated November 4, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Medtronic MiniMed Inc.
Location Northridge, CA
Product Type Devices
Quantity US 12,400,098 and Internationally 8,266,733

Product Description

Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets; Paradigm Silhouette infusion sets: MMT-368, MMT-368600, MMT-371, MMT-373, MMT-377, MMT-377600, MMT378, MMT-378600, MMT-38 1, MMT-381 600, MMT-382, MMT-382600, MMT-383, MMT-383600, MMT384, & MMT384600. Sure-T infusion sets MMT-862, MMT-863, MMT-864, MMT-865, MMT-866, MMT-873, MMT-874, MMTMMT-875, MMT-876, MMT-884, & MMT-886. MiniMed Duo sensor/infusion sets: MMT-5001A23, MMT-5001A43, MMT-5001B23, MMT-5001B43, MMT-5011A23, MMT-5011A43, MMT-5011B23, & MMT-5011B43. Silhouette and Sure-T disposable infusion sets are used with Medtronic insulin infusion pumps using Paradigm P-cap connectors. The infusion set connects the insulin reservoir in the pump to the patients body.

Reason for Recall

Tubing may detach at the connect/disconnect location on the MiniMed Silhouette and MiniMed Sure-T infusion sets. Detachment causes interrupted insulin delivery and the pump will not alarm to notify your patient. This can lead to hyperglycemia, and diabetic ketoacidosis (DKA).

Distribution Pattern

Worldwide distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lativa, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Venezuela.

Lot / Code Information

This field correction involves all models of Silhouette and Sure-T insulin infusion sets.

Other Recalls from Medtronic MiniMed Inc.

Recall # Classification Product Date
Z-1501-2016 Class II CareLink iPro2 Reports, Model No. MMT-7340 is ... Feb 19, 2016
Z-0438-2016 Class II MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model ... Sep 25, 2015
Z-0437-2016 Class II MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 ... Sep 25, 2015
Z-2442-2015 Class II Guardian Real-Time Monitor, Model No. CSS7100. ... Jul 22, 2015
Z-2260-2015 Class II MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-... Jun 19, 2015

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.