Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The...

FDA Device Recall #Z-0224-2015 — Class II — October 24, 2012

Recall Summary

Recall Number Z-0224-2015
Classification Class II — Moderate risk
Date Initiated October 24, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Nidek Inc
Location Fremont, CA
Product Type Devices
Quantity 21affected devices

Product Description

Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases.

Reason for Recall

When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.

Distribution Pattern

US Distribution in the states of: NJ, SC, IL, MI, NY, CA, TX, OH, IN, UT, and MN.

Lot / Code Information

Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,

Other Recalls from Nidek Inc

Recall # Classification Product Date
Z-1494-2017 Class II Final Fit Software Version 1.11 and 1.12; PC ... Feb 1, 2017
Z-2711-2016 Class II SPECULAR MICROSCOPE CEM-530; Software version 1... Aug 8, 2016
Z-2587-2016 Class II OPTICAL BIOMETER AL-Scan: Software Version: 1.0... Jul 1, 2016
Z-1245-2016 Class II RT-5100 (Epic 5100) Refractor and RT 3100 Refra... Feb 29, 2016
Z-2477-2015 Class II OPD-Scan III Refractive Power/Corneal Analyzer ... Jun 15, 2015

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.