Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

FDA Recall #Z-0582-2025 — Class II — October 31, 2024

Recall #Z-0582-2025 Date: October 31, 2024 Classification: Class II Status: Ongoing

Product Description

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

Reason for Recall

Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.

Recalling Firm

Wilson-Cook Medical Inc. — Winston Salem, NC

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

46 devices

Distribution

Distribution was OUS only to Argentina, Brazil, and Israel.

Code Information

UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated