Ingenia 1.5T S Model Number (REF): 781347

FDA Device Recall #Z-1847-2024 — Class II — May 3, 2024

Recall Summary

Recall Number Z-1847-2024
Classification Class II — Moderate risk
Date Initiated May 3, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips North America Llc
Location Cambridge, MA
Product Type Devices
Quantity 139 units

Product Description

Ingenia 1.5T S Model Number (REF): 781347

Reason for Recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Lot / Code Information

UDI-DI: (01)00884838068421(21) Serial Numbers: 81003 81050 81051 81090 81111 81010 81027 81028 81030 81035 81063 81106 81093 81105 81116 81007 81004 81006 81009 81011 81016 81017 81018 81019 81020 81023 81024 81031 81037 81041 81055 81056 81057 81060 81062 81064 81070 81074 81078 81112 81115 81122 81146 81147 81153 81156 81158 81013 81044 81040 81047 81059 81167 81170 81171 81008 81108 81148 81133 81164 81001 81086 81079 81118 81138 81139 81140 81151 81175 81184 81069 81061 81065 81098 81085 81091 81092 81032 81043 81129 81131 81149 81155 81162 81165 81172 81180 81084 81087 81088 81094 81095 81100 81127 81132 81174 81157 81176 81071 81052 81067 81039 81038 81179 81185 81117 81173 81154 81025 81097 81152 81015 81029 81033 81034 81042 81046 81048 81049 81058 81072 81080 81103 81113 81123 81128 81137 81045 81075 81076 81077 81096 81114 81124 81126 81145 81159 81178

Other Recalls from Philips North America Llc

Recall # Classification Product Date
Z-1316-2026 Class II IQon Spectral CT; Product Code (REF): 728332; Jan 7, 2026
Z-1318-2026 Class II Spectral CT 7500 on Rails; Product Code (REF):... Jan 7, 2026
Z-1317-2026 Class II Spectral CT; Product Code (REF): 728333; Jan 7, 2026
Z-1315-2026 Class II Brilliance iCT; Product Code (REF): 728306; Jan 7, 2026
Z-1033-2026 Class II Philips Smart-hopping 2.0 AP 1.4 GHz. Patient M... Dec 12, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.