3mensio Workstation (Vascular Fenestrated) software
FDA Recall #Z-2297-2025 — Class II — July 14, 2025
Product Description
3mensio Workstation (Vascular Fenestrated) software
Reason for Recall
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
Recalling Firm
PIE Medical Imaging B.V. — Maastricht, N/A
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
501
Distribution
US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.
Code Information
Software Version/UDI/Software Release Date: 10.6/(01)08056304455505(11)240325(8012)10.6/05-Apr-24, 10.7/(01)08056304455925(11)250423(8012)10.7/29-Apr-25, Service Packs/UDI: 10.6 SP1/(01)08056304455505(11)240411(8012)10.6 SP1/23-Apr-24, 10.6 SP2/(01)08056304455505(11)240610(8012)10.6 SP2/17-Jun-24, 10.6 SP3/(01)08056304455505(11)240926(8012)10.6 SP3/03-Oct-24, 10.6 SP4/(01)08056304455505(11)250109(8012)10.6 SP4/20-Jan-25, 10.7 SP1/(01)08056304455925(11)250523(8012)10.7 SP1/28-May-25
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.