NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01R...
FDA Recall #Z-1623-2025 — Class II — March 10, 2025
Product Description
NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, REF: NIM4CM01RF; PATIENT INTERFACE NIM4CPB1 NIM 4.0, REF: NIM4CPB1; PATIENT INTFC NIM4CPB1RF NIM 4.0 REFURB, REF: NIM4CPB1RF; SOFTWARE NIM4SWU143 UPGRADE V1.4.3, REF: NIM4SWU143; SOFTWARE NIM4SWU154 UPGRADE V1.5.4, REF: NIM4SWU154; SOFTWARE NIM4SWU164 UPGRADE V1.6.4, REF: NIM4SWU164
Reason for Recall
Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.
Recalling Firm
Medtronic Xomed, Inc. — Jacksonville, FL
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
9,788
Distribution
Worldwide - US Nationwide distribution including in the states of MD, MN, PA, OK, OH, FL, CO, GA, NC, KS, IL, WY, AK, NE, VA, MI, CA, NY, WI, UT, AZ, AR, TX, WA, AL, NJ, SD, MS, KY, TN, MO, OR, MT, MA, CT, NM, LA, WV, NH, DC, DE, IN, NV, IA, SC, ID, ME, HI, RI, VT, ND, GU, PR and the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Namibia, Nepal, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
Code Information
All NIM Vital Nerve Monitoring Systems manufactured or installed with software version v1.6.4 or earlier. REF/UDI-DI: NIM4CM01/763000002978, 00763000395896, 00763000401597, 00763000528577; NIM4CM01RF/00763000002992; NIM4CPB1/763000002985, 00763000401603, 00763000395902, 00763000528584; NIM4CPB1RF/00763000003005; NIM4SWU143/00763000709341, 00763000869823; NIM4SWU154/00763000945398; NIM4SWU164/00763000974312
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated