Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fix...
FDA Device Recall #Z-0639-2024 — Class II — October 16, 2023
Recall Summary
| Recall Number | Z-0639-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 16, 2023 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Zimmer Biomet Spine Inc. |
| Location | Westminster, CO |
| Product Type | Devices |
| Quantity | 3 |
Product Description
Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System
Reason for Recall
Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Distribution Pattern
US Nationwide distribution in the states of MN, NE.
Lot / Code Information
UDI-DI: 00889024002159, Lot: W834061
Other Recalls from Zimmer Biomet Spine Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0868-2019 | Class II | Vitality¿ Spinal Fixation System.Spinal Interve... | Jul 30, 2018 |
| Z-0866-2019 | Class II | Vitality¿ Spinal Fixation System. Thoracolumbos... | Jul 30, 2018 |
| Z-0867-2019 | Class II | Vitality¿ Spinal Fixation System. Spinal Interl... | Jul 30, 2018 |
| Z-2318-2017 | Class II | Vitality Spinal Fixation System, Vitality T27 F... | May 10, 2017 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.