Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/C...

FDA Recall #Z-1310-2026 — Class II — November 17, 2025

Recall #Z-1310-2026 Date: November 17, 2025 Classification: Class II Status: Ongoing

Product Description

Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.

Reason for Recall

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Recalling Firm

ETAC A/S — Gedved, N/A

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

64 units

Distribution

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

Code Information

(1) Model/Catalog Number: 1530101; UDI-DI: TBD; (2) Model/Catalog Number: 1530102; UDI-DI: TBD;

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.

Back to All Device Recalls