Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Prot...

FDA Recall #Z-0131-2026 — Class II — September 4, 2025

Recall #Z-0131-2026 Date: September 4, 2025 Classification: Class II Status: Ongoing

Product Description

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40

Reason for Recall

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.

Recalling Firm

Contego Medical LLC — Raleigh, NC

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

568

Distribution

US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.

Code Information

REF/UDI-DI/Lot(Expiration): NG-0740-140-2/00851616007613/Z2460707D(Z2460707D), Z2460733A(03/05/2027), Z2460766A(04/07/2027), Z2460774B(04/17/2027), Z2460780A(04/18/2027); NG-0640-140-2/00851616007590/Z2460746C(03/20/2027), Z2460774A(04/18/2027); NG-NV-7-40/00851616007620/Z2460776D(04/14/2027); NG-NV-6-40/00851616007651/Z2460776B(04/16/2027)

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated