AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Model/Catalog Number: E6...

FDA Recall #Z-0819-2025 — Class II — November 26, 2024

Recall #Z-0819-2025 Date: November 26, 2024 Classification: Class II Status: Ongoing

Product Description

AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Model/Catalog Number: E6-1222

Reason for Recall

The extension sets subject to this recall have ENFit connections, although they should have non- ENFit connections. ENFit connectors will only fit with other ENFit-compatible devices, and will not fit with non-ENFit connectors.

Recalling Firm

Applied Medical Technology Inc — Brecksville, OH

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

100

Distribution

US Distribution to states of: CA, CT, FL, GA, IL, MA, MD, MO, NC, NJ, NY, OH, OK, PA, SC, TX, VA, WA

Code Information

Lot Code: Model No: E6-1222 UDI-DI: (01)00842071107086(17)270401(10)240424-628 Lot number: 240424-628 Exp. Date: 2027-04-01

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated