Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Numbe...

FDA Recall #Z-0415-2026 — Class II — October 1, 2025

Recall #Z-0415-2026 Date: October 1, 2025 Classification: Class II Status: Ongoing

Product Description

Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control

Reason for Recall

The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.

Recalling Firm

Microgenics Corporation — Fremont, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

4462

Distribution

Worldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI, IL, LA, AR, TX, UT, GA, ID and the countries of MEX, CAN, KOR, MYS, DEU, GBR, BRA, JAP, HKG, NZL, AUS.

Code Information

OCRD-UL: UDI/DI 00884883011946, Lot Number OCRD2704U, Exp 2027-04-30; OCRD-L: UDI/DI 00884883011953, Lot Number OCRD2704L, Exp 2027-04-30; OCRD-101: UDI/DI 00884883011960, Lot Number OCRD27041, Exp 2027-04-30; OCRD-202: UDI/DI 00884883011977, Lot Number OCRD27042, Exp 2027-04-30; OCRD-303: UDI/DI 00884883011984, Lot Number OCRD27043, Exp 2027-04-30; OCRD-SP: UDI/DI 00884883011991, Lot Number OCRD2704S, Exp 2027-04-30; OCRD-MP: UDI/DI 00884883012004, Lot Number OCRD2704M, Exp 2027-04-30.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated