BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD st...

FDA Device Recall #Z-0026-2025 — Class II — September 12, 2024

Recall Summary

Recall Number Z-0026-2025
Classification Class II — Moderate risk
Date Initiated September 12, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Becton Dickinson & Co.
Location Sparks, MD
Product Type Devices
Quantity 162 units

Product Description

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place. Catalog Number: 444150

Reason for Recall

Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy

Distribution Pattern

Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan, Japan, Austria, Belgium, Denmark, France, Germany, Italy, Luxembourg, Netherlands, Norway, Spain, Sweden, Switzerland, United Kingdom, Russia, Saudi Arabia, South Africa, Canada, Mexico, China

Lot / Code Information

UDI-DI: 00382904441500 Software versions: 4.20, 4.30, 5.10, 5.20 and 5.30

Other Recalls from Becton Dickinson & Co.

Recall # Classification Product Date
Z-0484-2026 Class II BD Kiestra" Isolate Suspension Cuvette Array; ... Oct 23, 2025
Z-0235-2026 Class II BD EpiCenter Microbiology Data Management Syste... Sep 23, 2025
Z-0236-2026 Class II BD BACTEC Blood Culture System, labeled as the ... Sep 23, 2025
Z-0240-2026 Class II BD Veritor Connect Software, labeled as the fol... Sep 23, 2025
Z-0237-2026 Class II BD BACTEC Blood Culture System, labeled as the ... Sep 23, 2025

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.