SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus
FDA Recall #Z-0771-2021 — Class II — November 24, 2020
Product Description
SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus
Reason for Recall
System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
Recalling Firm
Siemens Medical Solutions USA, Inc — Malvern, PA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
3044
Distribution
US Nationwide
Code Information
SOMATOM Force (Model #10742326) SOMATOM Definition As (Model #8098027) SOMATOM Definition Edge (Model #10590000) SOMATOM Definition Flash (Model #10430603) SOMATOM Drive (Model #10431700) SOMATOM Confidence (Model #10590100) SOMATOM Edge Plus (Model #10267000) Siemens SOMATOM CT scanner systems with software syngo.CT VB20A_SP2.
Status
Terminated
Voluntary / Mandated
FDA Mandated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.