Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgi...
FDA Device Recall #Z-1697-2024 — Class II — January 25, 2024
Recall Summary
| Recall Number | Z-1697-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 25, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Covidien, LP |
| Location | North Haven, CT |
| Product Type | Devices |
| Quantity | 36 total, OUS distribution only |
Product Description
Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation and/or ligation.
Reason for Recall
One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.
Distribution Pattern
International distribution in the countries of Portugal and Italy.
Lot / Code Information
UDI-DI 20884521086422 Lot D2F2101RY
Other Recalls from Covidien, LP
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2281-2026 | Class II | Covidien EndoStitch" Polysorb Single Use Loadin... | Apr 15, 2026 |
| Z-0485-2026 | Class II | Covidien Signia" Small Diameter Curved Tip Inte... | Sep 25, 2025 |
| Z-0486-2026 | Class II | Covidien Signia" Small Diameter Curved Tip Inte... | Sep 25, 2025 |
| Z-1891-2024 | Class II | Tri-Staple 2.0 Black Reinforced Intelligent Rel... | Apr 15, 2024 |
| Z-1470-2024 | Class II | Covidien Auto Suture" Blunt Tip Trocar, Product... | Feb 28, 2024 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.