HORIBA custom configured fluorescence instrument, modular Fluorolog-QM
FDA Device Recall #Z-2166-2024 — Class II — March 28, 2024
Recall Summary
| Recall Number | Z-2166-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 28, 2024 |
| Status | Ongoing |
| Voluntary | FDA Mandated |
Recalling Firm
| Firm | Horiba Instruments Incorporated |
| Location | Piscataway, NJ |
| Product Type | Devices |
| Quantity | 4 units |
Product Description
HORIBA custom configured fluorescence instrument, modular Fluorolog-QM
Reason for Recall
HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
Distribution Pattern
US Nationwide and Worldwide Distribution
Lot / Code Information
Modular Fluorolog-QM
Other Recalls from Horiba Instruments Incorporated
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0119-2020 | Class II | Clinical Chemistry, ABX Pentra Lipase CP, ABX ... | Oct 30, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.