Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the samp...

FDA Recall #Z-0496-2025 — Class II — October 21, 2024

Recall #Z-0496-2025 Date: October 21, 2024 Classification: Class II Status: Ongoing

Product Description

Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each processing step. Siemens Material Number(SMN): 11099326 All lots starting with N15 or 15 are expected to have the defect.

Reason for Recall

A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.

Recalling Firm

Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

9729 units

Distribution

US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Congo, Republic C¿te dIvoire Croatia Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kenya Kuwait Latvia Libya Malaysia Mexico Moldova, Rep. of Morocco Nepal Netherlands Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam

Code Information

Lot # UDI N1527024 (01)00630414596143(10)N1527024(17)20291002; N1526424 (01)00630414596143(10)N1526424(17)20291001; N1525924 (01)00630414596143(10)N1525924(17)20290924 ; N1525324 (01)00630414596143(10)N1525324(17)20290918 ; N1524224 (01)00630414596143(10)N1524224(17)20290911; N1523724 (01)00630414596143(10)N1523724(17)20290904; N1523024 (01)00630414596143(10)N1523024(17)20290827; N1521824 (01)00630414596143(10)N1521824(17)20290814; N1521424 (01)00630414596143(10)N1521424(17)20290808; N1519324 (01)00630414596143(10)N1519324(17)20290719; N1518924 (01)00630414596143(10)N1518924(17)20290716; N1518224 (01)00630414596143(10)N1518224(17)20290710; N1518324 (01)00630414596143(10)N1518324(17)20290710; N1516224 (01)00630414596143(10)N1516224(17)20290621; N1515624 (01)00630414596143(10)N1515624(17)20290612; N1514924 (01)00630414596143(10)N1514924(17)20290605; N1514124 (01)00630414596143(10)N1514124(17)20290530; N1513324 (01)00630414596143(10)N1513324(17)20290522; N1512824 (01)00630414596143(10)N1512824(17)20290515; N1512424 (01)00630414596143(10)N1512424(17)20290510; N1511224 (01)00630414596143(10)N1511224(17)20290506; N1510724 (01)00630414596143(10)N1510724(17)20290423; N1510124 (01)00630414596143(10)N1510124(17)20290419.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated