stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103...

FDA Recall #Z-1651-2025 — Class II — March 3, 2025

Recall #Z-1651-2025 Date: March 3, 2025 Classification: Class II Status: Ongoing

Product Description

stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103; Prosthesis, Elbow, Semi-Constrained, Cemented

Reason for Recall

A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092, Humeral Stem Trial 150mm Medium / Left, Red color indicator (correct)

Recalling Firm

Tornier, Inc — Bloomington, MN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

5 units

Distribution

OUS (International) Distribution to countries: Sweden, Australia, United Kingdom, Colombia, Germany

Code Information

GTIN 00846832015054, Batch Number CC0421265

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.