ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary interna...

FDA Recall #Z-1715-2024 — Class II — March 20, 2024

Recall #Z-1715-2024 Date: March 20, 2024 Classification: Class II Status: Ongoing

Product Description

ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 7-2357-018-04

Reason for Recall

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Recalling Firm

Zimmer, Inc. — Warsaw, IN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Distribution

Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.

Code Information

UDI: (01)00889024088566(17)290409(10)64325922 (01)00889024088566(17)290414(10)64325923 (01)00889024088566(17)290515(10)64396817 (01)00889024088566(17)290517(10)64396818 (01)00889024088566(17)290831(10)64464677 (01)00889024088566(17)300325(10)64714915 (01)00889024088566(17)300330(10)64714916 (01)00889024088566(17)300321(10)64714917 (01)00889024088566(17)320329(10)65324701 (01)00889024088566(17)320510(10)65387353 (01)00889024088566(17)330220(10)65716463 (01)00889024088566(17)330127(10)65716464 (01)00889024088566(17)330426(10)65982902 (01)00889024088566(17)330426(10)66010705 Lot Numbers: 64325922 64325923 64396817 64396818 64464677 64714915 64714916 64714917 65324701 65387353 65716463 65716464 65982902 66010705

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated