Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of ped...

FDA Device Recall #Z-0295-2025 — Class I — October 7, 2024

Recall Summary

Recall Number Z-0295-2025
Classification Class I — Serious risk
Date Initiated October 7, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips Respironics, Inc.
Location Murrysville, PA
Product Type Devices
Quantity 1,322 units

Product Description

Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

Reason for Recall

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.

Lot / Code Information

Model No. LD2110X23B; UDI-DI: 606959057432; All Serial Numbers.

Other Recalls from Philips Respironics, Inc.

Recall # Classification Product Date
Z-1617-2026 Class I Philips Respironics Trilogy Evo Universal, Soft... Mar 2, 2026
Z-1615-2026 Class I Philips Respironics Trilogy Evo O2, Software Ve... Mar 2, 2026
Z-1616-2026 Class I Philips Respironics Trilogy EV300, Software Ver... Mar 2, 2026
Z-1614-2026 Class I Philips Respironics Trilogy Evo, Software Versi... Mar 2, 2026
Z-0481-2026 Class II SimplyGo Mini, Standard Battery Kit; Product c... Oct 7, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.