Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET U...

FDA Recall #Z-0448-2026 — Class I — October 10, 2025

Recall #Z-0448-2026 Date: October 10, 2025 Classification: Class I Status: Ongoing

Product Description

Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING

Reason for Recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Recalling Firm

Medline Industries, LP — Northfield, IL

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

96,406 total units

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Code Information

HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486; ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070; SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708; UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated