Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PA...

FDA Device Recall #Z-0827-2024 — Class II — November 29, 2023

Recall Summary

Recall Number Z-0827-2024
Classification Class II — Moderate risk
Date Initiated November 29, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm American Contract Systems, Inc.
Location Temple Terrace, FL
Product Type Devices
Quantity 2717 units

Product Description

Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Distribution Pattern

US Nationwide distribution.

Lot / Code Information

a) AGNE20J, UDI/DI 00191072183187, Lot Numbers: 934231, exp. 10/30/2023; 900231, exp. 4/10/2024; 990231, exp. 1/10/2024; 687221, exp. 11/9/2023; 720221, exp. 10/7/2023; 757221, exp. 8/31/2023; 870221, exp. 5/10/2023; 702221, exp. 10/25/2023; 795221, exp. 7/24/2023; b) LLCR99, UDI/DI 00191072162755, Lot Numbers: 756221, exp. 9/1/2023; 720221, exp. 10/7/2023; 654221, exp. 12/12/2023; 949231, exp. 2/20/2024; c) LLLC26, UDI/DI 00191072171900, Lot Numbers: 749221, exp. 9/8/2023; 695221, exp. 11/1/2023; 997231, exp. 1/3/2024; 920231, exp. 3/21/2024; d) MMCR11J, UDI/DI none, Lot Numbers: 744221, exp. 9/13/2023; 939231, exp. 3/2/2024; 876231, exp. 5/4/2024; e) MMNS20H, UDI/DI none, Lot Numbers: 963231, exp. 2/6/2024; 959231, exp. 2/10/2024; f) OSIN77, UDI/DI 00191072170705, Lot Numbers: 780221, exp. 8/8/2023; 728221, exp. 9/29/2023; 668221, exp. 11/28/2023; 991231, exp. 1/9/2024, 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; UDI/DI 00191072190031, Lot Numbers: 912231, exp. 3/29/2024; 903231, exp. 4/7/2024; g) SANE32H, UDI/DI none, Lot Numbers: 838221, exp. 6/11/2023; 889221, exp. 4/21/2023; h) SANE32I, UDI/DI none, Lot Numbers: 673221, exp. 11/23/2023; 730221, exp. 9/27/2023; 784221, exp. 8/4/2023; 911231, exp. 3/30/2024; 969231, exp. 1/31/2024; i) UICR10U, UDI/DI none, Lot Numbers: 990231, exp. 1/10/2024; 921231, exp. 3/20/2024; 885231, exp. 4/25/2024; j) UICR33AW, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 931231, exp. 3/10/2024; 876231, exp. 5/4/2024; k) UILM77AR, UDI/DI 00191072139184, Lot Numbers: 953231, exp. 2/16/2024; 905231, exp. 4/5/2024; l) WENE21D, UDI/DI 00191072167958, Lot Numbers: 839221, exp. 6/10/2023; m) WENE21F, UDI/DI 00191072167958, Lot Numbers: 819221, exp. 6/30/2023; n) WENE21H, UDI/DI 00191072179326, Lot Numbers: 729221, exp. 9/28/2023; 686221, exp. 11/10/2023; o) WENE21H-01, UDI/DI 00191072179326, Lot Numbers: 980231, exp. 1/20/2024; p) WENI56F, UDI/DI 00191072169341, Lot Numbers: 799221, exp. 7/20/2023; q) WENI56H, UDI/DI 00191072179333, Lot Numbers: 700221, exp. 10/27/2023; r) WENI56H-01, UDI/DI 00191072179333, Lot Numbers: 652221, exp. 12/14/2023; 996231, exp. 1/4/2024

Other Recalls from American Contract Systems, Inc.

Recall # Classification Product Date
Z-1916-2025 Class II LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREM... Apr 21, 2025
Z-1426-2025 Class II Brand Name: Banner University Med Ctr. Produ... Jan 31, 2025
Z-1425-2025 Class II Brand Name: Banner Health Systems Product Name... Jan 31, 2025
Z-1424-2025 Class II Brand Name: Banner Health Systems Product Name... Jan 31, 2025
Z-3244-2024 Class II Pacemaker, COPM11B; Medical convenience kit Aug 7, 2024

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.