PREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal ...
FDA Recall #Z-0732-2019 — Class II — December 17, 2018
Product Description
PREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal Antigen enzyme immunoassay (EIA) is a screening or a semi-quantitative test system for the detection of capsular polysaccharide antigens of Cryptococcus neoformans in serum and cerebrosoinal fluid (CSF).
Reason for Recall
An enzyme reagent included in the kit is not maintaining stability through claimed product expiration, resulting in the potential for false negative results when tested with patient specimens and the positive control.
Recalling Firm
Meridian Bioscience Inc — Cincinnati, OH
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
184
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, CA, CO, DE, IL, KS, KY, LA, MD, MO, NC, ND, NJ, NY, SC, TX, and VA. The products were distributed to the following foreign countries: India and Italy.
Code Information
Lot Numbers: 602096K089, 602096K090, 602096K091
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?