REF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - P...

FDA Recall #Z-0873-2021 — Class II — December 16, 2020

Recall #Z-0873-2021 Date: December 16, 2020 Classification: Class II Status: Ongoing

Product Description

REF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.

Reason for Recall

Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction

Recalling Firm

Medtronic Neurosurgery — Irvine, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

N/A

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IL, KS, KY, LA, MA, MI, MO, NV, NM, NY, ND, OH, OK, TN, VA and the countries of Albania, Australia, Brazil, Cyprus, Ireland, Mexico, Qatar, Romania, Russian Federation, Serbia, South Africa, Spain, United Kingdom.

Code Information

Lot # 0219948893, 0220166870, 0220236215, 0220704352, 0220720015, 0220752505

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated