Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 11 MM, 83 MM Long Item Number: 1...
FDA Device Recall #Z-0674-2022 — Class II — January 7, 2022
Recall Summary
| Recall Number | Z-0674-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 7, 2022 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Biomet, Inc. |
| Location | Warsaw, IN |
| Product Type | Devices |
| Quantity | 11 units |
Product Description
Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 11 MM, 83 MM Long Item Number: 113631
Reason for Recall
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.
Lot / Code Information
Lot Number: 821860 UDI: (01)00880304462687(17) 290522(10)821860
Other Recalls from Biomet, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0212-2026 | Class II | BIOMET ZipTight, Acute AC Joint Implant, Single... | Sep 23, 2025 |
| Z-2619-2025 | Class II | Juggerknot Mini Soft Anchors, Model/Catalog Num... | Sep 9, 2025 |
| Z-0477-2025 | Class II | JuggerLoc/ToggleLoc 2.9mm Drill Guide, Model nu... | Oct 21, 2024 |
| Z-2179-2024 | Class II | RingLoc Bi-Polar Hip System Acetabular Cup, 41 ... | May 17, 2024 |
| Z-1607-2024 | Class II | Vanguard Knee System AS Tibial Bearing / DCM Ar... | Mar 13, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.