NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the...
FDA Recall #Z-1000-2022 — Class II — February 18, 2022
Product Description
NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets other than SARS-CoV-2 are not impacted by this Urgent Medical Device Correction Ref. Number: 100100
Reason for Recall
Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients
Recalling Firm
NeuMoDx Molecular Inc — Ann Arbor, MI
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
773 (72 US and 701 ROW) cases (48 cartridges/case)
Distribution
Worldwide distribution - US Nationwide distribution in the states of OH, MO, VA, MI, MI, FL, MD, IA, NY, MO and the countries of Italy, Germany, Switzerland, Sweden, Netherlands, Finland, Spain, United Kingdom, Belgium, Austria, Hong Kong, France, Denmark, Slovenia.
Code Information
Lot Numbers: 115424, 115425, 115426, 115427, 115429, 115431. GTIN: 10814278020274
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated