Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180045

FDA Recall #Z-1864-2023 — Class I — May 25, 2023

Recall #Z-1864-2023 Date: May 25, 2023 Classification: Class I Status: Ongoing

Product Description

Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180045

Reason for Recall

Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.

Recalling Firm

TELEFLEX LLC — Morrisville, NC

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

4155 units

Distribution

US Nationwide distribution including Puerto Rico.

Code Information

UDI/DI 14026704340467, Batch Numbers: 18GT26, 18KG10, 18LG06, 19AG11, 19AG21, 19CT69, 19DT06, 19FT29, 19FT49, 19GT43, 19HT66, 19JT55, 19KT17, 19LT08, 19LT32, 20AT49, 20BT53, 20DT15, 20ET04, 20GG27, KME20H2636, KME20K1812, KME20K2573, KME21L2619, KME21M1773, KME22E2900, KME22G0347, KME22J3224, KME22K0489

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated