PrisMax System, Product Code 955626
FDA Device Recall #Z-0991-2022 — Class II — March 15, 2022
Recall Summary
| Recall Number | Z-0991-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 15, 2022 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Baxter Healthcare Corporation |
| Location | Deerfield, IL |
| Product Type | Devices |
| Quantity | 1253 units |
Product Description
PrisMax System, Product Code 955626
Reason for Recall
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Lot / Code Information
GTIN 00085412639499, All Serial Numbers
Other Recalls from Baxter Healthcare Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1573-2026 | Class II | Brand Name: Welch Allyn, Inc. Product Name: We... | Feb 18, 2026 |
| Z-1368-2026 | Class II | Brand Name: Spectrum IQ Product Name: Infusion... | Jan 13, 2026 |
| Z-1121-2026 | Class II | SIGMA Spectrum Infusion Pump, Model/Catalog Num... | Nov 28, 2025 |
| Z-0847-2026 | Class II | Welch Allyn INF BAG, Thigh 1-Tube, Model/Catalo... | Oct 28, 2025 |
| Z-0233-2026 | Class II | Oral Probe (Product code 02893-000), accessory ... | Sep 17, 2025 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.