CardioTek EP-TRACER Software V2.x.
FDA Device Recall #Z-1283-2022 — Class II — April 25, 2022
Recall Summary
| Recall Number | Z-1283-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 25, 2022 |
| Status | Completed |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | CardioTek BV |
| Location | Maastricht-Airport |
| Product Type | Devices |
| Quantity | 146 units |
Product Description
CardioTek EP-TRACER Software V2.x.
Reason for Recall
Device did not pass electrical safety testing for adequate insulation.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of CA, NY, NC and the countries of South Korea, Mexico, Kazakhstan, Netherlands, Spain, Germany, Finland, Poland, Israel, Turkey, Chile, Ecuador, Austria, Croatia, Sweden, United Kingdom, Colombia, Italy, Portugal, Canada, Saudi Arabia, Belgium, Slovenia, Brazil, Hungary, Malaysia, Uzbekistan.
Lot / Code Information
EP-TRACER 38 UDI/DI: 04260441455418; Serial Numbers: 2019-1-01, 2019-1-02, 2019-1-03, 2019-1-04, 2019-1-05, 2019-1-06, 2019-2-01, 2019-2-02, 2019-2-03, 2019-2-04, 2019-2-05, 2019-3-01, 2019-3-02, 2019-3-03, 2019-3-04, 2019-3-05, 2019-3-06, 2019-3-07, 2019-3-08, 2019-3-09, 2019-3-10, 2019-4-01, 2019-4-02, 2019-4-03, 2019-4-04, 2019-4-05, 2019-4-06, 2019-4-07, 2019-4-08, 2019-4-09, 2019-4-10, 2019-5-01, 2019-5-02, 2019-5-03, 2019-5-04, 2019-5-05, 2020-1-01, 2020-1-02, 2020-1-03, 2020-1-04, 2020-1-05, 2020-1-06, 2020-1-07, 2020-1-08, 2020-1-09, 2020-1-10, 2020-1-11, 2020-1-12, 2020-1-13, 2020-1-14, 2020-1-15, 2020-2-01, 2020-2-02, 2020-2-03, 2020-2-04, 2020-2-05, 2020-2-06, 2020-2-07, 2020-2-08, 2020-2-09, 2020-2-10, 2020-2-11, 2020-2-13, 2020-2-14, 2020-2-15, 2020-4-01, 2020-4-02, 2020-4-03, 2020-4-04, 2021-1-01, 2021-1-02, 2021-1-03, 2021-1-04, 2021-1-05, 2021-1-06, 2021-1-07, 2021-1-08, 2021-2-01, 2021-2-02, 2021-2-03, 2021-2-04, 2021-2-06, 2021-2-07, 2021-2-08, 2021-2-09, 2021-2-10, 2021-2-11, 2021-2-12, 2021-2-13, 2021-2-14, 2021-2-15, 2021-3-01, 2021-3-02, 2021-3-03, 2021-3-04, 2021-3-05, 2021-3-06, 2021-3-07, 2021-3-08, 2021-3-09, 2021-3-10, 2021-3-11, 2021-3-12, 2021-3-13, 2021-3-14, 2021-3-15, 2021-5-01, 2021-5-02, 2021-5-03, 2021-5-04, 2021-5-05, 2021-5-06, 2022-1-01, 2022-1-03, 2022-1-04, 2022-1-05, 2022-1-06; EP-TRACER 102 UDI/DI: 04260441455234; Serial Numbers: 2019-1-11, 2019-2-11, 2019-2-12, 2019-2-13, 2019-2-14, 2019-2-15, 2019-2-16, 2019-4-21, 2019-4-22, 2019-4-23, 2019-4-24, 2019-4-25, 2019-4-26, 2019-4-27, 2020-1-26, 2020-1-27, 2020-1-28, 2020-1-29, 2020-1-30, 2020-2-26, 2020-2-27, 2020-2-28, 2020-2-29, 2020-2-30, 2021-1-13, 2021-1-14, 2021-2-27, 2022-1-26, 2022-1-27
Other Recalls from CardioTek BV
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2657-2017 | Class II | CardioTek EP-TRACER Software V1.x and V2.0 T... | Apr 13, 2017 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.