PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure cu...
FDA Device Recall #Z-0819-2022 — Class II — February 7, 2022
Recall Summary
| Recall Number | Z-0819-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 7, 2022 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | B Braun Medical Inc |
| Location | Breinigsville, PA |
| Product Type | Devices |
| Quantity | 6 units |
Product Description
PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure customized, tailor-made solutions Catalog Number: 8713030U
Reason for Recall
Units not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy, overinfusion, or underinfusion.
Distribution Pattern
US Distribution to states of: MA, VT
Lot / Code Information
Serial Numbers: 633299, 633307, 633310, 633313, 633315, 633318 GUIDID: 04046963716745
Other Recalls from B Braun Medical Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1481-2026 | Class II | Brand Name: B. Braun Product Name: BBraun Medi... | Feb 2, 2026 |
| Z-1482-2026 | Class II | Brand Name: B. Braun Product Name: 21GA WINGED... | Feb 2, 2026 |
| Z-1483-2026 | Class II | Brand Name: B. Braun Product Name: 21GA WINGED... | Feb 2, 2026 |
| Z-0656-2026 | Class II | IV Administration Set utilized in gravity IV ad... | Oct 29, 2025 |
| Z-0708-2026 | Class II | Outlook set utilized in gravity IV administrati... | Oct 29, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.