CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023...
FDA Device Recall #Z-1265-2022 — Class II — May 10, 2022
Recall Summary
| Recall Number | Z-1265-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 10, 2022 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare, LLC |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 30,993 Central Stations and 2,180 FRUs and displays |
Product Description
CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices: (1) Part #2082294-001, FRU MP200 POWER SUPPLY ASSEMBLY; (2) Part #2082293-002-R, FRU MP200 CPU ASSEMBLY; (3) Part #2082293-002, FRU MP200 CPU ASSEMBLY; (4) Part #2071523-001, MP200 22 IN DISPLAY WITH TOUCH; (5) Part #2092930-001, DSP LCD 22 IN MED W/TCH W/DKST N AMR; (6) Part #2092930-002, DSP LCD 22 IN MED W/TCH W/DKST BRIT P/C; (7) Part #2092930-003, DSP LCD 22IN MED W/TCH W/DKST EURO P/C; (8) Part #2092930-004, DSP LCD 22IN MED W/TCH W/DKST ITA P/C; (9), Part #2092930-005, DSP LCD 22IN MED W/TCH W/DKST INDIAN P/C; (10) Part #2092930-006, DSP LCD 22IN MED W/TCH W/DKST ISREAL P/C; (11) Part #2092930-007, DSP LCD 22IN MED W/TCH W/DKST SWISS P/C; (12) Part #2092930-008, DSP LCD 22IN MED W/TCH W/DKST AUSTRL P/C; (13) Part #2092930-009, DSP LCD 22IN MED W/TCH W/DKST JPN; (14) Part #2092930-010, DSP LCD 22IN MED W/TCH W/DKST BRZL P/C; (15) Part #2092930-011, DSP LCD 22IN MED W/TCH W DKST ARGNT P C; (16) Part #2092930-012, DSP LCD 22IN MED W TCH W DKST DENMARK; (17) Part #2092930-013, DSP LCD22IN MED W TCH W DKST S AFRICA P C; and (18) Part #2092930-014, DSP LCD22IN MED W TCH W DKST CHINA P C.
Reason for Recall
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
Distribution Pattern
Worldwide distribution - US Nationwide. There was also military/government distribution and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Kazakhstan, Kenya, Korea, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Malta, Mauritius, Mexico, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, and Yemen.
Lot / Code Information
All CARESCAPE Central Station V2 units: (1) CSCS V2 MA1700 Integrated, Product Code SKN, Model #2082278-001, GTIN 00840682109666; (2) CSCS V2 MAS700 Desktop, Product Code SNF, Model #2082279-001, GTIN 00840682109604. NOTE: The Product Code is the first 3 letters of the serial number.
Other Recalls from GE Healthcare, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2022 | Class II | (1) GE Centricity Universal Viewer 6.0, Model 2... | Jul 15, 2022 |
| Z-1430-2022 | Class II | SIGNA Premier magnetic resonance scanner, model... | Jun 24, 2022 |
| Z-1466-2022 | Class II | Centricity PACS Software Version 7.0 SP0.0.4.7 | Jun 14, 2022 |
| Z-1333-2022 | Class II | GE Centricity Universal Viewer Zero Footprint. ... | Jun 8, 2022 |
| Z-1511-2022 | Class II | Service manuals for the CT, PET, or NM table as... | Jun 2, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.