CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for art...
FDA Recall #Z-1294-2023 — Class II — February 27, 2023
Product Description
CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip arthroplasty. Item Number: 01.01012.384
Reason for Recall
Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009
Recalling Firm
Zimmer, Inc. — Warsaw, IN
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
1,168 untits
Distribution
AL AZ CO FL LA MI OH OK TX Foreign: Canada
Code Information
GTIN Number: 00889024283268 All lots
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated