AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850
FDA Recall #Z-0692-2022 — Class II — December 8, 2021
Product Description
AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850
Reason for Recall
Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
Recalling Firm
Beckman Coulter Inc. — Brea, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
181,776 Kits
Distribution
US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming; and OUS: Albania, Algeria, American Samoa, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Ethiopia, Finland ,France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Lithuania, Macau, Macedonia, Malawi, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Niger, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saint Pierre and Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Vietnam, Yemen, Zambia, Zimbabwe.
Code Information
All lot numbers.
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated