Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System

FDA Recall #Z-1049-2023 — Class II — December 16, 2022

Recall #Z-1049-2023 Date: December 16, 2022 Classification: Class II Status: Ongoing

Product Description

Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System

Reason for Recall

Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.

Recalling Firm

NovaSignal Corp. — Los Angeles, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

6

Distribution

US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL

Code Information

UDI-DI/Serial Numbers: 00850681007221/ 21020003, 20080004, 20100004, 21080001, 20120001; 00850681007009/ 19090005

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated