Compress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotom...

FDA Recall #Z-0371-2024 — Class II — October 9, 2023

Recall #Z-0371-2024 Date: October 9, 2023 Classification: Class II Status: Ongoing

Product Description

Compress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178402

Reason for Recall

Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery

Recalling Firm

Biomet, Inc. — Warsaw, IN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

N/A

Distribution

Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, CANADA, CHILE, JAPAN, NETHERLANDS, NEW ZEALAND.

Code Information

Lot Numbers/UDI 240750 (01)00880304259713(17)270112(10)240750 671330 (01)00880304259713(17)271114(10)671330 671420 (01)00880304259713(17)270831(10)671420 671660 (01)00880304259713(17)270901(10)671660 752730 (01)00880304259713(17)270802(10)752730 752760 (01)00880304259713(17)270816(10)752760 798010 (01)00880304259713(17)270120(10)798010 921660 (01)00880304259713(17)270724(10)921660 930670 (01)00880304259713(17)270727(10)930670 930680 (01)00880304259713(17)270802(10)930680 965330 (01)00880304259713(17)271112(10)965330

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.

Back to All Device Recalls