GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Us...
FDA Recall #Z-1101-2021 — Class II — December 10, 2020
Product Description
GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts.
Reason for Recall
GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
Recalling Firm
GE Healthcare, LLC — Waukesha, WI
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
113 units
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV
Code Information
Serial Numbers: G1D1842032, G1D1842035, G1C17X2022, G1D1852039, Not Available*, G1D1922084, G1B16X2002, G1D1972106, G1D18X2058, G1D18Y2066, G1B1912074, G1B1772014, G1C1782016, G1D1842033, G1D18Z2071, G1D1962102, G1D1962104, G1D1862041, G1D18Y2068, G1D1972107, G1D18Z2069, Not Available*, G1D19X2118, G1D1982112, Not Available*, G1D1862043, G1D18Y2065, Not Available*, G1D1962101, G1C17X2026, G1D1832029, G1D1962105, G1B1772012, G1D1832030, G1B1772012, G1D1852038, Not Available, G1D1842031, G1C17X2023, G1C1782017, G1D1892054, G1D1932088, G1C17X2024, G1D1882049, G1D1882048, G1D1912077, G1D1932089, GID194092, G1923001923, G1793810017, G1D1882052, G1D1922080, G1D1862045, G1D1882051, G1C1792020, G1D1892055, G1D1932090, G1B1742006, G1D18Z2070, G1D1862042, G1D18X2062, G1D18X2061, G1B1762011, G1B1762010, G1D1942093, G1D1862044, G1D1882050, G1B1912079, G1D1992113, G1B1762009, G1D1852036, G1D1992117, G1D1842034, G1D1922083, G1D1952097, G1D18Z2073, G1D18X2060, G1D1912078, G1D1922085, G1D1892057, Not Available*, G1D1932091, G1D1922081, Not Available*, G1D1912075, G1D18Z2072, G1D1852040, G1D1892056, G1C1792021, G1D1982111, G1D1952100, G1B16X2005, G1D1952096, G1D1962103, G1D1922082, G1D1972108, Not Available*, G1D1932086, G1B16X2004, G1D1852037, G1D18X2059, G1D18Y2064, G1D18Y2063, G1B1942095, G1D1932087, G1D1913076, G1D18Y2067, G1D1872046, G1D1832028, G1D1992114, G1C1792018, G1C17X2025, G1B16X2003 not available* - to be supplied
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated