Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9....
FDA Device Recall #Z-1281-2023 — Class II — February 17, 2023
Recall Summary
| Recall Number | Z-1281-2023 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 17, 2023 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | RAYSEARCH LABORATORIES AB |
| Location | Stockholm, N/A |
| Product Type | Devices |
| Quantity | 10770 systems |
Product Description
Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs
Reason for Recall
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
Distribution Pattern
US Nationwide
Lot / Code Information
UDI-DI: 0735000201067920221007;
Other Recalls from RAYSEARCH LABORATORIES AB
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1110-2026 | Class II | Brand Name: RayStation Product Name: RayStatio... | Nov 28, 2025 |
| Z-1106-2026 | Class II | Brand Name: RayStation Product Name: RayStatio... | Nov 28, 2025 |
| Z-1109-2026 | Class II | Brand Name: RayStation Product Name: RayStatio... | Nov 28, 2025 |
| Z-1104-2026 | Class II | RayStation/RayPlan. Model/Catalog Numbers: 12.0... | Nov 28, 2025 |
| Z-1108-2026 | Class II | Brand Name: RayStation Product Name: RayStatio... | Nov 28, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.