Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10....
FDA Recall #Z-0878-2013 — Class II — January 15, 2010
Product Description
Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty.
Reason for Recall
Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.
Recalling Firm
Medacta Usa Inc — Camarillo, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
293 units
Distribution
Nationwide Distribution including the states of ID, CO, TX, WI, NY, CT, WA, CA, UT, and AZ.
Code Information
Reference Number Lot Number Reference Number Lot Number 01.26.10.0133: 095920, 096050, 096050A; 01.26.10.0138: 095925, 096055, 096055A; 01.26.10.0134: 095921, 096051, 096051A; 01.26.10.0139: 095926, 096056, 096056A; 01.26.10.0135: 095922, 096052, 096052A; 01.26.10.0140: 095927, 096057, 096057A; 01.26.10.0136: 095923, 096053, 096053A; 01.26.10.0141: 095928, 096058, 096058A; 01.26.10.0137: 095924, 096054, 096054A; 01.26.10.0142: 095929, 096059, 096059A.
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated