Philips Healthcare Brilliance iCT Computed Tomography X-Ray System

FDA Device Recall #Z-1714-2016 — Class II — September 6, 2015

Recall Summary

Recall Number Z-1714-2016
Classification Class II — Moderate risk
Date Initiated September 6, 2015
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips Healthcare
Location Andover, MA
Product Type Devices
Quantity US-423

Product Description

Philips Healthcare Brilliance iCT Computed Tomography X-Ray System

Reason for Recall

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Distribution Pattern

US Distribution - AZ, PA, FL, DC, W VA, NC, SC, AR, OK, OH, NY, NV, TX, PA, IN, CA, MN, MA, DC, MS, MD, OR, IL, CO, MO, VT, IA, AL, LA, GA and WA.

Lot / Code Information

Software v. 4.13, 4.14 and 4.15.

Other Recalls from Philips Healthcare

Recall # Classification Product Date
Z-2520-2023 Class II Incisive CT, software version 5.0 Aug 2, 2023
Z-0737-2022 Class II Philips StentBoost Live R2.0 application, when ... Feb 2, 2022
Z-0238-2022 Class II Philips Azurion systems with software releases ... Sep 24, 2021
Z-0241-2022 Class II CombiDiagnost R90 is multi-functional general R... Sep 22, 2021
Z-0239-2022 Class II DigitalDiagnost C90 Sep 22, 2021

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.