Recall Summary
| Recall Number | Z-1563-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 4, 2020 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | LivaNova USA Inc |
| Location | Houston, TX |
| Product Type | Devices |
| Quantity | 2683 |
Product Description
VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators.
Reason for Recall
False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.
Distribution Pattern
U.S.: LA, ID, AZ, UT, KY, VA, TN, NY, SC, IL, IN, NC, WA, CA, OH, FL, PA, NJ, CO, TX, RI, MD, HI, OR, MA, NE, MO, OK, MS, MN, SD, VT, AR, WV, WI, MI, AL, KS, DE, NM, GA, ND, AK, MT, ME, NV, DC, MO, WY, IA, CT. Canada, and Israel.
Lot / Code Information
Model 3000 version 1.6
Other Recalls from LivaNova USA Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0535-2021 | Class II | VNS Therapy SENTIVA DUO # 1000-D - Product Usag... | Nov 12, 2020 |
| Z-0462-2021 | Class II | VNS Therapy¿ AspireSR¿ Generator | Sep 28, 2020 |
| Z-0561-2020 | Class I | VNS Therapy, SenTiva, Model # 1000, (01)0542502... | Aug 22, 2019 |
| Z-1118-2020 | Class II | VNS Therapy SENTIVA Generator Model # 1000 ... | Aug 13, 2019 |
| Z-0663-2020 | Class II | VNS Therapy Program GMDN: 44077 Model # 3000 V1... | Nov 16, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.