Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conju...
FDA Recall #Z-0835-2021 — Class II — November 23, 2020
Product Description
Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Reason for Recall
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
Recalling Firm
Fresenius Medical Care Holdings, Inc. — Waltham, MA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
1426 cases
Distribution
US Nationwide distribution.
Code Information
Lot Numbers: 20NXAC013 20NXAC035 Additional Affected lots as of 3/17/21: 20LXAC056 20LXAC041 20LXAC055 20LXAC057 20NXAC013 20NXAC035
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.