CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnosti...

FDA Recall #Z-1329-2020 — Class II — December 19, 2019

Recall #Z-1329-2020 Date: December 19, 2019 Classification: Class II Status: Ongoing

Product Description

CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.

Reason for Recall

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Recalling Firm

Philips Medical Systems Gmbh, DMC — Hamburg

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

131

Distribution

Worldwide distribution  US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Code Information

(domestic only) SN374 SN17081704 SN17121782 SN18011800 SN18081862 SN18061846 SN18111907

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated