TroClose 1200 - Product Usage: intended for use in a variety of gynecologic, general and urologic...

FDA Device Recall #Z-0398-2021 — Class II — September 22, 2020

Recall Summary

Recall Number Z-0398-2021
Classification Class II — Moderate risk
Date Initiated September 22, 2020
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Gordian Surgical
Location Misgav, N/A
Product Type Devices
Quantity 300 total devices

Product Description

TroClose 1200 - Product Usage: intended for use in a variety of gynecologic, general and urologic endoscopic procedures, to create and maintain a port of entry and to facilitate the delivery of absorbable sutures and anchors through soft tissues of the body during endoscopic / laparoscopic surgery. The trocar may be used with or without visualization for primary and secondary insertions. It is comprised of a single patient use, bladeless Obturator designed to allow penetration and positioning at the required site within the abdomen, and several single use Cannulas, which serve as working channels for the procedure. The Cannulas are pre-loaded with two synthetic absorbable sutures, each attached to a synthetic absorbable anchor. The deployment of the anchors and sutures is achieved by utilizing a mechanism within the Obturator. The Cannula is 115 mm long and is compatible for use with laparoscopic devices of 5 - 12 mm outer diameter.

Reason for Recall

Endotoxin test results found above the acceptable levels in LAL test (above 20EU/device).

Distribution Pattern

US Nationwide distribution including in the states of Texas and Michigan.

Lot / Code Information

Catalog Number: GSTC1200 Lot Numbers: AS200009/3, Exp 14 FEB 2022 AS200012/4, Exp 23 MAR 2022

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.