Vios Monitoring System Bedside Monitor Model BSM2050

FDA Device Recall #Z-1292-2025 — Class II — August 9, 2024

Recall Summary

Recall Number Z-1292-2025
Classification Class II — Moderate risk
Date Initiated August 9, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Murata Vios, Inc.
Location Woodbury, MN
Product Type Devices
Quantity 50

Product Description

Vios Monitoring System Bedside Monitor Model BSM2050

Reason for Recall

Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.

Distribution Pattern

US Nationwide distribution in the states of Florida, Louisiana, Maryland, Montana, New York, Ohio, and Texas.

Lot / Code Information

1.56.01 UDI: +B595BSM20500/$$7V1.56.01/16D20230630S 1.57.20 UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y Software Versions 1.56.01 and 1.57.20

Other Recalls from Murata Vios, Inc.

Recall # Classification Product Date
Z-1069-2025 Class II muRata Vios Monitoring System Model 2050; bedsi... Apr 22, 2024
Z-1169-2025 Class II Vios Monitoring System Bedside Monitor, Model B... Jun 6, 2023
Z-0956-2025 Class II muRata Vios Monitoring System Model 2050; bedsi... Jun 21, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.