EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light s...

FDA Device Recall #Z-0870-2020 — Class II — September 27, 2019

Recall Summary

Recall Number Z-0870-2020
Classification Class II — Moderate risk
Date Initiated September 27, 2019
Status Ongoing
Voluntary FDA Mandated

Recalling Firm

Firm Energetiq Technology Inc
Location Woburn, MA
Product Type Devices
Quantity 15

Product Description

EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.

Reason for Recall

On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.

Distribution Pattern

Worldwide Distribution.

Lot / Code Information

The defect concerns a total of fifteen (15) of our EQ-400 products. manufactured between June 3, 2016 and September 26, 2019.

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.