CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory ...
FDA Device Recall #Z-0885-2020 — Class II — October 21, 2019
Recall Summary
| Recall Number | Z-0885-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 21, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Haemonetics Corporation |
| Location | Braintree, MA |
| Product Type | Devices |
| Quantity | 44821 |
Product Description
CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
Reason for Recall
Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
Lot / Code Information
Catalog number/FG Item Number: 00260-00 1. Specific Lots with Bowl Cracks (6920 Distributed) Lot numbers: 1907006HTT 1902005HTT 1902008HTT 1902011HTT 1906005HTT 1803013HTT 1804004HTT 1804006HTT 1804002HTT 1804009HTT 1805002HTT 1805003HTT 1811002HTT 1812010HTT 812011HTT 1901001HTT 1812005HTT 1812008HTT 1901007HTT 1808006HTT 1808010HTT 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
Other Recalls from Haemonetics Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1310-2022 | Class II | TEG 5000 Functional Fibrinogen Reagent, Item No... | Apr 26, 2022 |
| Z-1298-2022 | Class II | Product Name: TEG5000 Analyzer (07-022, 07-033)... | Apr 19, 2022 |
| Z-1617-2020 | Class II | TEG Manager software versions used in conjuncti... | Feb 19, 2020 |
| Z-0890-2020 | Class II | CS5/5+ Fastpacks, 225 mL, 20¿ Res Autotransfusi... | Oct 21, 2019 |
| Z-0900-2020 | Class II | JAPAN DISP SET, 261J, 205J, HAR - Product Usage... | Oct 21, 2019 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.