Single Use Distal Cover MAJ-2315 Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed...
FDA Recall #Z-1071-2024 — Class II — December 8, 2023
Product Description
Single Use Distal Cover MAJ-2315 Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed to be attached to Olympus duodenovideoscopes to cover the distal end of the insertion tube and around the forceps elevator
Reason for Recall
Single Use Distal Cover MAJ-2315 may unexpectedly detach from the duodenoscope during use result include the risk of mucosal injury, tissue damage, bleeding, perforation, or obstruction due to retained distal cover in the gastrointestinal tract that would require urgent medical intervention for removal, and/or burns from the use an uncovered duodenoscope distal end. Detachment of a distal cover in the oral cavity may result in aspiration, airway obstruction, or respiratory distress and would require immediate medical intervention for removal.
Recalling Firm
Olympus Corporation of the Americas — Center Valley, PA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
24300 units (box of 20 pieces)
Distribution
US Nationwide distribution.
Code Information
UDI: 04953170441271 All lot numbers
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated