Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Mo...

FDA Recall #Z-0857-2021 — Class II — December 10, 2020

Recall #Z-0857-2021 Date: December 10, 2020 Classification: Class II Status: Ongoing

Product Description

Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.

Reason for Recall

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Recalling Firm

GE Healthcare, LLC — Waukesha, WI

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

52 untits

Distribution

Worldwide

Code Information

Discovery 670 DR (Model H3100BT), Serial Numbers: DRGY56033, DRGY56037, DRGY56035, DRGY56030, DRGZ56043, DRGZ56042, DRGY56032, DRGY56031, DRGY56034, DRGY56036 NM/CT 870 DR (Model 3100Q), Serial Numbers: 870Y64066, and (Model H3100AS), Serial Numbers: 870Y64077, 870Y64064, 870Y64081, 870Y64079, 870Y64067, 870Y64074, 870Z64117, 870Y64078, 870Y64080, 870Y64069, 870Y64085, 870Z64121, 870Z64119, 870Z64122, 870Y64056, 870Y64063, 870Z64114, 870Z64115, 870Y64068, 870Y64082, 870Z64116, 870Z64120, 870Y64065, 870Y64084, 870Y64061, 870Y64083, 870Z64113, 870Y64070, 870Z64123, 870Y64058, 870Z64118, 870Y64057 Discovery NM670 Pro (Model 3100XB), Serial Numbers: PRGY52027, PRGY52031, PRGZ52047, PRGZ52046, PRGY52028, PRGZ52048, PRGY52029, PRGY52030 Discovery 670 ES (Model H3100XF), Serial Numbers: ESGZ53005

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.